1. Multi-Parameter Monitoring: Measures NIBP, SpO₂, PR, temperature, and optional mainstream CO₂ for comprehensive care.
2. High Precision NIBP: Offers species-specific pressure ranges and an accuracy of ±5 mmHg / ±3%.
3. Ultra-Long Battery Life: Delivers up to 15 hours of continuous operation.
4. Flexible & Compact Design: Portable 3.5" screen with versatile handheld, desktop, or wall-mount use.
5. Advanced Data & Reliability: Stores 256 animal profiles + 30,000 records, with anti-motion and low-perfusion performance.
| Category | Specification | Details |
| Physical | Dimensions | 175 mm (h) × 85 mm (w) × 32 mm (d) |
| Weight | NIBP+SpO₂+TEMP: 1348.5g SpO₂ only: 1171.5 g SpO₂+TEMP: 1220g |
| Power | External Power | 100–240 VAC, 0.4A, 50/60 Hz, ± 10% |
| Battery Type | Rechargeable Lithium-ion |
| Battery Capacity | 3.7V / 4000 mAh |
| Battery Life | ≥ 15 hours |
| Display | Type | 3.5" Color LCD Touchscreen |
| Resolution | 480 × 854 |
| Environmental | Application Range | Dogs, Cats, Horse |
| Model Configurations | MT-VM-002-A | NIBP + SpO₂ + TEMP |
| MT-VM-002-B | SpO₂ only |
| MT-VM-002-C | SpO₂ + TEMP |
| SpO₂ | Measurement Accuracy | ±2% (70% – 100%), undefined (0 – 69%) |
| Measurement Range | 0–100% |
| Temperature (TEMP) | Measurement Accuracy | ±0.2°C (0°C – 50°C) |
| Measurement Range | 0°C – 50°C |
| Minimum Time to Valid Reading | Approx. 3 minutes |
| NIBP (Blood Pressure) | Measurement Method | Oscillometric |
| Operation Modes | Manual, Auto, Continuous |
| Systolic Range (Horse/Dog) | 30 mmHg – 270 mmHg |
| Systolic Range (Cat) | 30 mmHg – 220 mmHg |
| Diastolic Range (Horse/Dog) | 10 mmHg – 220 mmHg |
| Diastolic Range (Cat) | 10 mmHg – 195 mmHg |
| Mean Pressure Range (Horse) | 20 mmHg – 235 mmHg |
| Mean Pressure Range (Dog) | 20 mmHg – 235 mmHg |
| Mean Pressure Range (Cat) | 20 mmHg – 205 mmHg |
| Accuracy | Mean error ≤ ±5 mmHg or ±3% (whichever is larger) |
| Cuff Pressure Range | 0 mmHg – 300 mmHg |
| Cuff Inflation Speed | ≤ 50 mmHg/s |
| Static Pressure Error | ± 3 mmHg (without overpressure protection) |
| Pulse Rate Range | 40 bpm – 240 bpm |
| Pulse Rate Accuracy | ±1% or ±2 bpm (whichever is larger) |
| Pulse Rate | Measurement Range | 20 bpm ~ 254 bpm |
| Alarm Range | Same as display range; alarm limit step: 1 bpm |
| Refresh Rate | Approx. 1 second |
| Accuracy | ±3 bpm |
| Averaging Period | 8 seconds |
| Pulse Tone | Non-modulated |
| Wavelength Range | 500 nm ~ 1000 nm (for all sensors) |
| Display Format | Numeric |
| Data Storage | Animal Profiles | 256 |
| Maximum Data Records | 30,000 |
| Trend Graph Display | Up to 120 hours |
| Continuous Data Logging | Up to 20 days |